Microplate Manager User Guide
84
Absorbance Report
Figure 57. Absorbance Report.
In the Absorbance Report, the mean absorbance of all the blank wells in an assay is
subtracted from the raw absorbance for each non-blank well in the same assay. A
negative value will be reported if the mean absorbance of the blank wells is greater
than the absorbance of the unknown well.
Absorbances are displayed only for standards, unknowns, and positive and negative
controls.
6.6.4 Reports Available for Kinetic Data Only
Velocity Report