Bio-rad Aurum™ Total RNA 96 Kit Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Bio-rad Aurum™ Total RNA 96 Kit. Bio-Rad Aurum™ Total RNA 96 Kit User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Total RNA 96 Kit

Catalog #732-6800Aurum™ Total RNA 96 Kit Instruction Manual

Strona 2

6 Disruption and Homogenization Proper disruption and homogenization of the starting materials are required to ensure complete lysis of the cells an

Strona 3 - Table of Contents

7 Preparing the Aurum™ Vacuum Manifold Tubing provided in the Aurum™ Vacuum Manifold kit is 4 ft long and must be cut into appropriate pieces before

Strona 4

8 2. Place the desired 96-well binding plate on the manifold top and apply the recommended vacuum pressure for your application. Manifold Elution

Strona 5 - Kit Components

9 Section 7 Vacuum Protocol PleasereadSection5,“BeforeUsingtheAurum™TotalRNA96Kit”andSection 6, “Vacuum Manifold Setup and Use With 96-

Strona 6 - Necessary Supplies

10 Yeast Follow steps C1–C5, then continue with step 1 of “All Starting Cell Types” onpage10.Ifstartingwithagrowblockofyeastculture(max

Strona 7 - Total RNA 96 Kit

11 5. Add700µloflowstringencywashsolutiontoeachwelloftheRNAbinding plate. Gradually increase the negative pressure between –17 to –23

Strona 8 - S. cerevisiae

12 Note: Gradual application of negative pressure is required to prevent sample spraying and cross-contamination. The eluted total RNA sample

Strona 9

13B2. Add350µloflysissolution(alreadysupplementedwith 1%b-mercaptoethanol) to each sample and pipet up and down several times to mix thoro

Strona 10 - 96-Well Plates

14 5. Add700µloflowstringencywashsolutiontoeachwelloftheRNAbinding plate. Centrifuge for 2 min at 1,500 x g. Discard the low stringen

Strona 11 - Vacuum manifold

15 Section 9 Troubleshooting Guide Problem Possible Cause Recommended Solution Difficulty achieving Purge step in protocol If sealing of well

Strona 13 - Vacuum Protocol

16 Problem Possible Cause Recommended Solution Low or highly variable Elution solution applied Apply elution solution eluate volumes among to

Strona 14

17 Problem Possible Cause Recommended Solution Clogging of RNA Excessive amount of Reduce volume of binding plate starting material per well

Strona 15

18 Problem Possible Cause Recommended Solution Total RNA prep Incorrect use of wash Add the appropriate performs poorly in  solutions volum

Strona 16 - Spin Protocol

19 Section 10 OrderingInformation Catalog # Description 732-6800 Aurum™ Total RNA 96 Kit 732-6470 Aurum™ Vacuum Manifold 732-6820 Aurum™

Strona 17

Bio-Rad Laboratories, Inc. 2000 Alfred Nobel Dr. Hercules, CA 94547 USA (510) 741-1000 1-800-424-6723 Life ScienceGroup10-1554 1210 Sig 111

Strona 18

Table of Contents Section 1 Introduction...1 Section 2 Kit Components ...1 Section 3 Storage

Strona 20

Section 1 Introduction The Aurum™ Total RNA 96 kit rapidly purifies up to 192 total RNA samples from biological samples (e.g. mammalian cells, yea

Strona 21

Section 3 Storage Conditions All kit components (including lyophilized DNase I) should be stored at room temperature. Store reconstituted DNase I

Strona 22

Section 5 BeforeUsingthe Aurum™ Total RNA 96 Kit Please read the following guidelines before proceeding with the total RNA purification. Start

Strona 23 - OrderingInformation

Table 1. Yield (per well) of total RNA from various samples using the Aurum™ Total RNA 96 kit. Starting Material Avg. Yield (µg)* Cultured cell

Strona 24 - 1-800-424-6723

• Vendors of lyticase, which is used to partially degrade the cell walls of yeast cells, may have different definitions of the enzyme’s activity.

Komentarze do niniejszej Instrukcji

Brak uwag