Step 3
Set Up the
High Voltage (HV)
Conditioning Routine
6
1. Insert a blank ProteinChip array into the reader. If blank arrays are not
available, use expired arrays or arrays that have already been used.
a. For the Enterprise Edition instrument: place a cassette filled with
12 blank or used arrays into the instrument.
b. For the Personal Edition instrument: insert a blank ProteinChip
array.
2. In ProteinChip data manager software, select the instrument and click Start.
3. In the Protocol Mode tab, select Instrument > Quick Run.
4. Select all 8 spots, then set up a Quick Run protocol as shown below:
a. For the Enterprise Edition instrument:
RC3 HV pos 15 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 20 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 30 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV neg 15 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV neg 20 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV neg 25 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV neg 30 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 15 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 20 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 30 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
b. For the Personal Edition instrument:
RC3 HV pos 15 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 20 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 30 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 15 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 20 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 30 kV conditioning, partition 1 of 8, all 8 spots
RC3 HV pos 25 kV conditioning, partition 1 of 8, all 8 spots
5. Click Save in the Quick Run dialog, and save the protocol as
SoftHV
in the
Quick Run directory (C:\\XXXXXXX). The Quick Run directory location is
defined by the user. The default location, "My Documents", will work as
well as any other location that is easy to remember.
Komentarze do niniejszej Instrukcji