Bio-rad ChromLab™ Software Instrukcja Użytkownika Strona 258

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 300
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 257
7 | Evaluating Results
256 | NGC Chromatography Systems and ChromLab Software
Displaying Column Performance Results — the
Column Performance Table
The results of the column performance analysis are detailed in the Column
Performance table. The table displays the following fields for the column
performance peak:
N — displays the calculated number of theoretical plates for the column.
N/L (cm
-1
) — displays the calculated number of theoretical plates per
meter for the column.
HETP (cm) — displays the calculated height equivalent to a theoretical
plate for the column.
rHETP — displays the calculated reduced plate height for the column.
Bed height (cm) — displays the value for medium bed height as specified
in the Settings pane.
Particle diameter (m) — displays the value for average particle diameter
as specified in the Settings pane.
Run start time
Column type — displays the type of column. The column type is set in the
Method Editor tab in the Method Settings phase.
Start — the beginning of the peak, displayed in the units of the currently
chosen x-axis.
End — the end of the peak, displayed in the units of the currently chosen
x-axis.
Retention time — displays the elution time at maximum peak height.
Retention volume — displays the elution volume at maximum peak
height.
Height (mAU) — displays the peak height at the retention volume/time,
measured from the UV trace to the calculated baseline.
Przeglądanie stron 257
1 2 ... 253 254 255 256 257 258 259 260 261 262 263 ... 299 300

Komentarze do niniejszej Instrukcji

Brak uwag