
Calibrations
User Guide | 6 5
Calibrating Flow Rate of Pumps
If the buffer flow seems to be incorrect, you might want to recalibrate the pumps.
Important: The system must be primed with water befor
e you calibrate the
pumps’ flow rate. Otherwise the calibration will be incorrect.
To calibrate the flow rate of the pumps
Select Pump Flow Rate in the Calibrate dropdown list at the top of the
Calibration dialog box.
The Calibration dialog box displays instructions and settings for calibrating the
pumps.
Komentarze do niniejszej Instrukcji