Verifying Plumbing with the Point-to-Plumb Feature
User Guide | 6 7
Verifying Plumbing with the Point-to-Plumb Feature
The NGC chromatography system instrument arrives plumbed. The Point-to-Plumb
feature enables you to change the plumbing or to verify that the instrument is
plumbed correctly for the fluidic scheme you want to use.
You can access the Point-to-Plumb feature on the touch screen menu or on the
computer running Chr
omLab. (Select T
ools > Point-to-Plumb.)
Tip: Using the touch screen makes it easy to view the Point-to-Plumb window
and the fr
ont of the instrument at the same time.
When you select Point-to-Plumb, the curr
ent fluidic scheme appears in
Point-to-Plumb mode. All LEDs on the instrument tur
n of
f. When you select a line
between two modules, the LEDs corresponding to the two ports to be connected
light up.
In the Point-to-Plumb window, the proper plumbing path is indicated by the color
gray. Selected flow paths turn green. Path elements that are not plumbed appear in
a paler shade of gray and do not change color when you select them.
Note: When the fluidic scheme includes a signal import module (SIM), the SIM
is also pr
esent in the Point-to-Plumb window
. In this case, the outlet LED on the
module that connects to the external detector (via SIM) turns on when the fluid
Przeglądanie stron 68