A | Maintaining the Instrument
214 | NGC Chromatography Systems and ChromLab Software
Replacing the Conductivity Monitor
The images in this section show the conductivity monitor on the single-wavelength
UV detector module. Though the position of the conductivity monitor differs on the
multi-wavelength UV/Vis module, the steps to replace the conductivity monitor are
the same.
To replace the conductivity monitor
1. On the touch screen, select Shut Down on the dropdown menu to exit
ChromLab and shut down the NGC instrument.
2. Disconnect the tubing lines from the conductivity monitor.
3. Loosen the two thumbscrews on the front of the detector.
Przeglądanie stron 215