D | Connecting the C-96 Autosampler to NGC Systems
290 | NGC Chromatography Systems and ChromLab Software
5. Locate the two cables that connect the autosampler to the SIM:
15-pin D connector cable (part #10027227)
9-pin D connector cable (part #10027224)
These cables ship with the autosampler.
6. Attach the DIGITAL IN connector to the 15-pin D connector cable.
Tip: The 15-pin D connector cable has three wires (red, black, and
br
own).
a. From left to right, loosen the first, second, and last screws on the
connector
.
b. If necessary, strip the plastic wire coating off the wires so that they are
exposed.
c. Insert the wires into the open ports in this order and tighten the screws:
Insert the red wire into the first port.
Insert the black wire into the second port.
Insert the brown wire into the last (ground) port.
Note: Take care not to overtighten or fray the wires.
7. Attach the DIGITAL OUT connector to the 9-pin D connector cable.
Tip: The 9-pin D connector cable has two wires (red and black).
a. From left to right, loosen the first and last screws on the connector.
b. Insert the wires into the open ports in this order and tighten the screws,
taking car
e not to overtighten or fray the wires:
Insert the red wire into the last port.
Insert the black wire into the first (ground) port.