5 | Plumbing and Priming the NGC Chromatography Systems
118 | NGC Chromatography Systems and ChromLab Software
Plumbing the NGC Discover Pro System
Table 12 lists the tasks for plumbing the NGC Discover Pro system. The tasks are
explained in detail in the sections that follow.
Note: Ensure the you complete the plumbing tasks in the sections Plumbing
All NGC Systems on page 100 and Plumbing the NGC Discover System on
page 113 before continuing.
1.
Tube #6 between the outlet port on the sample inlet valve and the
inlet port on the sample pump
2.
Connect one of the following:
Tube #4 from the Common inlet port on the BioFrac fraction
collector diverter valve to port 1 on the outlet valve
Waste tubing to port 1 on the outlet valve
3.
Inlet tubing to the remaining ports on the sample inlet valve
See the section Plumbing Sample Inlet Valves on page 120.
4.
Outlet tubing to the remaining ports on the outlet valve
See the section Plumbing Outlet Valves on page 121.
The figure on page 119 shows the Point-to-Plumb connections for the
NGC Discover Pro system.
Table 12. Plumbing tasks for the NGC Discovery Pro systems
Tasks
Note: Remove the inlet tubing from the sample pump inlet
manifold if it is installed.