Priming and Purging the System
Installation Guide | 123
Priming and Purging the System
This section explains how to prime your system before running the pumps, and
how to purge your system of air bubbles or rapidly exchanging buffer
solutions.
Note: This section assumes that you primed the wash system when you
plumbed the system. See Plumbing All NGC Systems on page 100 for more
information.
Priming and purging tasks are specific to the NGC system. Table 13 lists the tasks
associated with each system configuration. The tasks are explained in detail in the
sections that follow
.
Prime and purge the
inlet port on system
pump A
Prime and purge Q1–Q4
on the buffer blending
valve
Prime and purge Q1–Q4 on
the buffer blending valve
Prime and purge the
inlet port on system
pump B
Prime and purge the buffer
inlet valves A and B
Prime and purge the
sample pump
Table 13. Priming and purging tasks for each NGC system configuration
NGC Quest NGC Scout NGC Discover
Tip: The inlet ports
on system pumps A
and B are primed
and purged when
the buffer blending
valve is primed.
Tip: The inlet ports on
system pumps A and B
ar
e primed and pur
ged
when the buffer
blending valve is
primed.