Priming and Purging the System
Installation Guide | 129
Purging the NGC Scout System
To purge the buffer blending valve
1. Open the System Pump dialog box and select Priming in the Mode dropdown
list.
To purge an individual line of air bubbles, proceed to step 2 on page 129.
To purge all lines of air bubbles sequentially, proceed to step 3 on
page 130.
2. To manually purge an individual line of air bubbles and fill it with buffer:
a. Select a port to open.
b. (Optional) Change the default flow rate.
The default flow rate is set to 20 ml/min.
c. Click Start.
d. Repeat steps 2a–c for any other port to purge.