Error Messages
Operating panel
Message Meaning Possible cause Check Remedy
Perform operating panel diagnostics
Battery dead Charge battery, check battery acid,
replace battery
Charge indicator lamp defective Check charge indicator lamp,
Check wiring
Dynamo defective Replace the dynamo
Internal error, operating panel Replace operating panel
Perform operating panel diagnostics
CU voltage 5V
electronics too low
Error: 5 Volt
Electronics voltage
too low
Error: 5 Volt
Electronics voltage
CU voltage 5V
electronics too
126
127
The controller of the dynamo is
defective
Replace the dynamo
Dynamo defective Replace the dynamo
Battery change-over relay defective Replace battery change-over relay
Internal error, operating panel Replace operating panel
Electronics voltage
too high
electronics too
high