Check Remedy
DIOM
N° Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
Check wiring to valve
16024 DIOM DIG3 status Error: Status
16023 DIOM DIG2 status Error: Status
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16025 DIOM DIG4 status Error: Status