Error Messages
CAN
Possible cause Check RemedyMeaningMessage
Power supply voltage
CU defective
LED -16-LD1 not lit Check fuse -22-F47
Check fuse -22-F48
Power supply voltage
cent
ral electrical system defective
LED -22-LD62 not lit Check fuse -22-F62
LED -22-LD64 not lit Check fuse -22-F64
Wiring faulty Check wiring and plug
Internal error, operating panel Replace operating panel
Power supply voltage
DIOM defective
LED -22-LD59 not lit Check fuse -22-F59
LED -22-LD27 not lit Check fuse -22-F73
Power supply voltage
central electrical system defective
LED -22-LD63 not lit Check fuse -22-F63
Error: Restart CU
har
dware
9961
9965 Restart DIOM
hardware
Error: Restart DIOM
hardware
Restart CU
hardware
central electrical system defective
LED -22-LD60 not lit Check fuse -22-F60
Check fuse -22-F92
Wiring faulty Check wiring and plug
Internal error DIOM Replace DIOM
Power supply voltage
ADM1 defective
LED -22-LD17 not lit Check fuse -22-F68
Check fuse -22-F70
Power supply voltage
cent
ral electrical system defective
LED -22-LD64 not lit Check fuse -22-F64
LED -22-LD63 not lit Check fuse -22-F63
Check fuse -22-F92
Wiring faulty Check wiring and plug
Internal error ADM1 Replace ADM1
Error: Restart ADM1
hardware
Restart ADM1
hardware
9966