BIO RAD 400-0174 Instrukcje Operacyjne Strona 454

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 485
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 453
Error Messages
Possible cause Check Remedy
KMC1
Message Meaning
Run electronics diagnostics
Power supply voltage
KMC1 too low
LED -22-LD61 not lit Check fuse -22-F61
Power supply voltage
ce
ntral electrical system defective
LED -22-LD60 not lit Check fuse -22-F60
Check fuse -22-F92
KMC1 LD36 not lit
Wiring defective Check wiring
Short circuit in the wiring
to a 12 V sensor
Check wiring
1101 KMC1 electrical
voltage too low
Error:
Electronics voltage
too low
to a 12 V sensor
Charge indicator lamp defective Check the charge indicator lamp, check
the wiring
The controller of the dynamo is
defective
Check the excitation voltage;
Check wiring
Dynamo defective Replace the dynamo
Battery dead Charge battery, check battery acid,
replace battery
Battery change-over relay defective Replace battery change-over relay
Internal error - KMC1 Replace KMC1
Przeglądanie stron 453
1 2 ... 449 450 451 452 453 454 455 456 457 458 459 ... 484 485

Komentarze do niniejszej Instrukcji

Brak uwag