Error Messages
Possible cause Check Remedy
KMC1
Message Meaning
Run electronics diagnostics
Output voltage of V2
defective
LED -22-LD38 not lit Check fuse -22-F86
Central electrical power supply voltage
defective
LED -22-LD60 not lit Check fuse -22-F60
LED -22-LD63 not lit Check fuse -22-F63
Check fuse -22-F92
Fuse F6 in KMC1 defective
KMC1 LD37 not lit
Check fuse F6 in KMC1
Wiring of release switch defective
Check wiring
1104 KMC1 voltage V2
defective!
Error: KMC1
Output voltage of V2
Wiring of release switch defective
Check wiring
Release switch defective Replace release switch
Check release logic, and
replace GAL component, if necessary
GAL component defective Replace GAL component
Charge indicator lamp defective Check charge indicator lamp,
Check wiring
Dynamo defective Check the excitation voltage, check the
wiring, replace the dynamo
Battery dead Charge battery, check battery acid,
replace battery
Internal error - KMC1 Replace KMC1
GAL component release logic wrong
(wrong GAL component)